1.
J Clin Monit Comput ; 32(1): 63-72, 2018 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-28074417

RESUMO

Two types of Planecta™ ports are commonly used as sampling ports in blood pressure transducer kits: a flat-type port (FTP) and a port with a three-way stopcock (PTS). Recently, a new type of three-way stopcock (Marvelous™) has been released as a Planecta™ counterpart, but its effects on the frequency characteristics and reliability of blood pressure monitoring have not been investigated. We assessed the influence of the Marvelous™ stopcock on the frequency characteristics of the pressure transducer kit. The basic pressure transducer kit, DT4812J, was modified by replacing one or two of the original three-way stopcocks with Marvelous™ stopcocks. The frequency characteristics (i.e., natural frequency and damping coefficient) of each kit were determined using wave parameter analysis software, and subsequently evaluated on a Gardner chart. Replacement of the original blood pressure transducer kit stopcocks with Marvelous™ stopcocks decreased the natural frequency (48.3 Hz) to 46.3 Hz or 44.8 Hz, respectively; the damping coefficient was not significantly changed. Plotting the data on a Gardner chart revealed that the changes fell within the adequate dynamic response region, indicating they were within the allowable range. Insertion of Marvelous™ stopcocks slightly affects the natural frequency of the pressure transducer kit, similar to inserting a PTS. The results indicate that the Marvelous™ stopcock is useful for accurate monitoring of arterial blood pressure, and may be recommended when insertion of two or more closed-loop blood sampling systems is necessary.


Assuntos
Determinação da Pressão Arterial/instrumentação , Determinação da Pressão Arterial/métodos , Transdutores de Pressão , Pressão Sanguínea , Desenho de Equipamento , Humanos , Oscilometria/instrumentação , Oscilometria/métodos , Reprodutibilidade dos Testes , Processamento de Sinais Assistido por Computador , Software
2.
J Clin Monit Comput ; 31(2): 371-380, 2017 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-26946147

RESUMO

The accurate monitoring of arterial blood pressure is important for cardiovascular management. However, the frequency characteristics of pressure transducer kits are influenced by the length of the pressure-resistant tube. To date, there have been few studies addressing the frequency characteristics of pressure transducer kits with inserted pressure-resistant extension tubes (pressure-resistant extension tube (ET) circuits). In this study, we examine ET circuits from the viewpoint of the frequency characteristics of pressure transducer kits. DT4812J transducer kits (length 150 cm; Argon Medical Devices, TX, USA) were used. Three original ET circuits were prepared, with the pressure-resistant tube of the DT4812J being extended with a 30-cm length of pressure-resistant tube (180ET circuit), a 60-cm length of pressure-resistant tube (210ET circuit), and a 90-cm length of pressure-resistant tube (240ET circuit). Each of these circuits was evaluated as part of this study. The natural frequency of the original DT4812J circuit was 45.90 Hz while the damping coefficient was 0.160. For the 180 ET circuit, the natural frequency and damping coefficient were 36.4 Hz and 0.162, respectively. For the ET210 circuit, the natural frequency and damping coefficient were 30.3 Hz and 0.175, respectively. For the ET210 circuit, the natural frequency and damping coefficient were 25.3 Hz and 0.180, respectively. As a result of extending the circuit, it was found that the natural frequency decreased drastically, while the damping coefficient increased slightly. When the extension of a pressure transducer kit is required, we should pay careful attention to the major decrease in the natural frequency, which may influence the pressure monitoring.


Assuntos
Determinação da Pressão Arterial/instrumentação , Transdutores de Pressão , Transdutores , Pressão Sanguínea , Determinação da Pressão Arterial/métodos , Desenho de Equipamento , Humanos , Modelos Estatísticos , Pressão , Reprodutibilidade dos Testes , Processamento de Sinais Assistido por Computador
3.
J Clin Monit Comput ; 30(6): 925-931, 2016 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-26467334

RESUMO

Blood pressure transducer kits are equipped with two types of Planecta™ ports-the flat-type Planecta™ port (FTP) and the Planecta™ port with a three-way stopcock (PTS). We reported that FTP application decreased the natural frequency of the kits. However, Planecta™ is an invaluable tool as it prevents infection, ensures technical simplicity, and excludes air. Hence, an ideal Planecta™ port that does not decrease the frequency characteristics is required. As a first step in this direction, we aimed to assess the influence of PTSs on the natural frequency of blood transducer kits. A DTXplus transducer kit (DT4812J; Argon Medical Devices, TX, USA) was used along with ≥1 PTSs (JMS, Hiroshima, Japan), and the frequency characteristics were assessed. The natural frequency and damping coefficient of each kit were obtained by using frequency characteristics analysis software, and these parameters were evaluated by plotting them on Gardner's chart. Regardless of whether one or two PTSs were inserted, the natural frequency of the kits only slightly decreased (from 42.5 to 41.1 Hz, when 2 PTSs were used). Thus, the frequency characteristics of the kits with PTSs were adequate for pressure monitoring. The insertion of ≥2 FTPs in pressure transducer kits should be avoided, as they markedly decrease the natural frequency and lead to underdamping. However, the effect of PTS insertion in pressure transducer kits on the frequency characteristics is minimal. Thus, we found that the use of PTS markedly improved the frequency characteristics as compared to the use of FTP.


Assuntos
Determinação da Pressão Arterial/instrumentação , Determinação da Pressão Arterial/métodos , Monitores de Pressão Arterial , Cateterismo Periférico/instrumentação , Cateterismo Periférico/métodos , Transdutores de Pressão , Desenho de Equipamento , Humanos , Japão
4.
J Clin Monit Comput ; 29(6): 681-9, 2015 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-25516163

RESUMO

Pressure-transducer kits have frequency characteristics such as natural frequency and damping coefficient, which affect the monitoring accuracy. The aim of the present study was to investigate the effect of planecta ports and a damping device (ROSE™, Argon Medical Devices, TX, USA) on the frequency characteristics of pressure-transducer kits. The FloTrac sensor kit (Edwards Lifesciences, CA, USA) and the DTXplus transducer kit (Argon Medical Devices) were prepared with planecta ports, and their frequency characteristics were tested with or without ROSE™. The natural frequency and damping coefficient of each kit were obtained using frequency characteristics analysis software and evaluated by plotting them on the Gardner's chart. By inserting a planecta port, the natural frequency markedly decreased in both the FloTrac sensor kit (from 40 to 22 Hz) and the DTXplus transducer kit (from 35 to 22 Hz). In both kits with one planecta port, the damping coefficient markedly increased by insertion of ROSE™ from 0.2 to 0.5, optimising frequency characteristics. In both kits with two planecta ports, however, the natural frequency decreased from 22 to 12 Hz. The damping coefficient increased from 0.2 to 0.8 by insertion of ROSE™; however, optimisation was not achieved even by ROSE™ insertion. Planecta ports decrease the natural frequency of the kit. ROSE™ is useful to optimise the frequency characteristics in the kits without or with one planecta port. However, optimisation is difficult with two or more planecta ports, even with the ROSE™ device.


Assuntos
Determinação da Pressão Arterial/instrumentação , Transdutores de Pressão , Monitores de Pressão Arterial , Desenho de Equipamento , Humanos , Sístole , Análise de Ondaletas